General specifications |
Signal type | Analog input |
Functional safety related parameters |
Safety Integrity Level (SIL) | SIL 2 |
Supply |
Connection | terminals 23+, 24- or power feed module/Power Rail |
Rated voltage | 20 ... 30 V DC |
Rated current | approx. 100 mA |
Power dissipation/power consumption | ≤ 2 W / 2.2 W |
Interface |
Programming interface | programming socket |
Input |
Connection side | field side |
Connection | terminals 1, 2, 3, 4, 6
|
RTD | Pt100, Pt500, Pt1000, Ni100, Ni1000 |
| Types of measuring | 2-, 3-, 4-wire technology |
| Lead resistance | max. 50 Ω |
| Measurement loop monitoring | sensor breakage, sensor short-circuit |
Thermocouples | type B, E, J, K, L, N, R, S, T (IEC 584-1: 1995) |
| Cold junction compensation | external and internal |
| Measurement loop monitoring | sensor breakage |
Potentiometer | 0.8 ... 20 kΩ |
| Types of measuring | 2-, 3-, 5-wire technology |
Voltage | 0 ... 10 V , 2 ... 10 V , 0 ... 1 V , -100 ... 100 mV |
Input resistance | ≥ 250 kΩ (0 ... 10 V) min. 1 MΩ (0 ... 1 V, -100 ... 100 mV) |
Measuring current | approx. 400 µA with resistance measuring sensor |
Output |
Connection side | control side |
Connection | output I: terminals 10, 11, 12 output II: terminals 16, 17, 18 output III: terminals 8+, 7- |
Output I, II | relay |
| Contact loading | 250 V AC / 2 A / cos φ ≥ 0.7 ; 40 DC / 2 A |
| Mechanical life | 5 x 107 switching cycles |
| Energized/De-energized delay | approx. 20 ms / approx. 20 ms |
Output III | Analog current output |
| Current range | 0 ... 20 mA or 4 ... 20 mA |
| Open loop voltage | max. 24 V DC |
| Load | max. 650 Ω |
| Fault signal | downscale I ≤ 3.6 mA, upscale I ≥ 21 mA (acc. NAMUR NE43) |
Collective error message | Power Rail |
Transfer characteristics |
Deviation | |
| Temperature effect | Input: 0.005 %/K (50 ppm) of span ; current output: 0.005 %/K (50 ppm) of span |
| RTD | max. 0.2 % of span |
| Thermocouples | max. 10µV deviation of CJC: ±0.8 K |
| Voltage | 0.1 % of span |
| Potentiometer | 0.1 % of span when < 5 kΩ 0.5 % of span when > 5 kΩ |
| Current output | max. 20 µA |
Sampling rate | approx. 700 ms |
Galvanic isolation |
Input/Other circuits | reinforced insulation according to IEC/EN 61010-1, rated insulation voltage 300 Veff |
Output I, II against eachother | reinforced insulation according to IEC/EN 61010-1, rated insulation voltage 300 Veff |
Output I, II/other circuits | reinforced insulation according to IEC/EN 61010-1, rated insulation voltage 300 Veff |
Output III/power supply and collective error | reinforced insulation according to IEC/EN 61010-1, rated insulation voltage 300 Veff |
Interface/power supply | reinforced insulation according to IEC/EN 61010-1, rated insulation voltage 300 Veff |
Indicators/settings |
Display elements | LEDs , display |
Control elements | Control panel |
Configuration | via operating buttons via PACTware |
Labeling | space for labeling at the front |
Directive conformity |
Electromagnetic compatibility | |
| Directive 2014/30/EU | EN 61326-1:2013 (industrial locations) |
Low voltage | |
| Directive 2014/35/EU | EN 61010-1:2010 |
Conformity |
Electromagnetic compatibility | NE 21:2007 |
Degree of protection | IEC 60529:2001 |
Ambient conditions |
Ambient temperature | -20 ... 60 °C (-4 ... 140 °F) |
Mechanical specifications |
Degree of protection | IP20 |
Connection | screw terminals |
Mass | 300 g |
Dimensions | 40 x 119 x 115 mm (1.6 x 4.7 x 4.5 inch) (W x H x D) , housing type C2 |
| Height | 119 mm |
| Width | 40 mm |
| Depth | 115 mm |
Mounting | on 35 mm DIN mounting rail acc. to EN 60715:2001 |
Data for application in connection with hazardous areas |
EU-type examination certificate | TÜV 03 ATEX 2140 |
| Marking | II (1)G [Ex ia Ga] IIC II (1)D [Ex ia Da] IIIC I (M1) [Ex ia Ma] I |
| Input | Ex ia |
Supply | |
| Maximum safe voltage | 40 V DC (Attention! The rated voltage can be lower.) |
Input | terminals 2, 6 (for active equipment) |
| Voltage | 13.1 V |
| Current | 8 mA |
| Power | 67 mW |
| Voltage | 29 V |
| Current | 11 mA |
| Power | 200 mW |
Inputs | terminals 1, 2, 3, 4, 6 (for passive equipment) |
| Voltage Uo | 13.1 V |
| Current Io | 21 mA |
| Power Po | 67 mW |
Analog output | |
| Maximum safe voltage | 40 V (Attention! The rated voltage can be lower.) |
Interface | |
| Maximum safe voltage | 40 V (Attention! The rated voltage can be lower.) , RS 232 |
Certificate | PF 08 CERT 1213 X |
| Marking | II 3G Ex nA nC IIC T4 Gc |
Output I, II | |
| Contact loading | 50 V AC/2 A/cos φ > 0.7; 40 V DC/1 A resistive load |
Galvanic isolation | |
| Input/Other circuits | safe electrical isolation acc. to IEC/EN 60079-11, voltage peak value 375 V |
Directive conformity | |
| Directive 2014/34/EU | EN IEC 60079-0:2018+AC:2020 , EN 60079-11:2012 , EN 60079-15:2010 |
International approvals |
IECEx approval | |
| IECEx certificate | IECEx TUN 09.0019 |
| IECEx marking | [Ex ia Ga] IIC , [Ex ia Da] IIIC , [Ex ia Ma] I |
General information |
Supplementary information | Observe the certificates, declarations of conformity, instruction manuals, and manuals where applicable. For information see www.pepperl-fuchs.com. |